          Abstract

          Objectives

          To objectively evaluate posterior capsular opacification (PCO) with RTVue-100 spectral domain-OCT and assess the agreement with the Pentacam system.

          Methods

          Sixty-seven eyes diagnosed with PCO were included. RTVue-100 SD-OCT was used to scan the IOL outline and PCO at horizontal and vertical meridians. PCO was also imaged with a Pentacam and slit-lamp photography system. With RTVue-100 SD-OCT, the PCO area, thickness, density, and objective scores were recorded and used to evaluate the severity of PCO at 3 mm and 5 mm diameter ranges of the IOL optic region. We assessed the correlation of visual acuity, PCO characteristics, and PCO scores. PCO scores acquired from RTVue-100 SD-OCT images were also compared with those from the Pentacam. Differences between pear-type and fibrosis-type PCOs were also compared using RTVue-100 SD-OCT cross-sectional images.

          Results

          The cross-sectional images of PCO acquired with RTVue-100 SD-OCT corresponded well to Pentacam and slit-lamp retroillumination images. IOL-posterior capsular space, area, thickness, and density of the proliferated and accumulated LECs could be clearly visualized and quantified with RTVue-100 SD-OCT. PCO scores were correlated with decreased visual acuity, which was in line with the outcomes using the Pentacam. Differences between the pear-type and fibrosis-type PCO were statistically significant; pear-type PCOs showed a wider and thicker opacification region with lower density compared with fibrosis-type PCOs.

          Conclusion

          RTVue-100 SD-OCT could be a powerful tool in PCO objective evaluation and classification. OCT could be used to visualize the morphology and outline of PCO. Thus, it could discriminate and quantify differences between different types of PCO. PCO scores seem to be a useful factor that could reliably reflect PCO severity.

          Related collections

          Most cited references32

          • Record: found
          • Abstract: found
          • Article: not found

          Posterior capsule opacification

          A complication of extracapsular cataract extraction with or without posterior chamber intraocular lens (PC-IOL) implantation is posterior capsule opacification. This condition is usually secondary to a proliferation and migration of residual lens epithelial cells. Opacification may be reduced by atraumatic surgery and thorough cortical clean-up. Clinical, pathological and experimental studies have shown that use of hydrodissection, the continuous curvilinear capsulorhexis and specific IOL designs may help reduce the incidence of this complication. Capsular-fixated, one-piece all-polymethylmethacrylate PC-IOLs with a C-shaped loop configuration and a posterior convexity of the optic are effective. Polymethylmethacrylate loops that retain "memory" create a symmetric, radial stretch on the posterior capsule after in-the-bag placement, leading to a more complete contact between the posterior surface of the IOL optic and the taut capsule. This may help form a barrier against central migration of epithelial cells into the visual axis. Various pharmacological and immunological methods are being investigated but conclusive data on these modalities are not yet available.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Prevention of posterior capsular opacification.

            Posterior capsular opacification (PCO) is a common complication of cataract surgery. The development of PCO is due to a combination of the processes of proliferation, migration, and transdifferentiation of residual lens epithelial cells (LECs) on the lens capsule. In the past decades, various forms of PCO prevention have been examined, including adjustments of techniques and intraocular lens materials, pharmacological treatments, and prevention by interfering with biological processes in LECs. The only method so far that seems effective is the implantation of an intraocular lens with sharp edged optics to mechanically prevent PCO formation. In this review, current knowledge of the prevention of PCO will be described. We illustrate the biological pathways underlying PCO formation and the various approaches to interfere with the biological processes to prevent PCO. In this type of prevention, the use of nanotechnological advances can play a role.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Optical coherence tomography angiography: Technical principles and clinical applications in ophthalmology

              Optical coherence tomography angiography (OCTA) is a functional extension of OCT that provides information on retinal and choroidal circulations without the need for dye injections. With the recent development of high-speed OCT systems and efficient algorithms, OCTA has become clinically feasible. In this review article, we discuss the technical principles of OCTA, including image processing and artifacts, and its clinical applications in ophthalmology. We summarize recent studies which qualitatively or quantitatively assess disease presentation, progression, and/or response to treatment.
                Bookmark

                Author and article information

                Contributors
                Journal
                J Ophthalmol
                J Ophthalmol
                JOPH
                Journal of Ophthalmology
                Hindawi
                2090-004X
                2090-0058
                2018
                23 December 2018
                : 2018
                : 5461784
                Affiliations
                1Clinical College of Ophthalmology, Tianjin Medical University, Tianjin 300020, China
                2Tianjin Eye Hospital, Tianjin 300020, China
                3Tianjin Key Laboratory of Ophthalmology and Visual Science, Tianjin, China
                4Tianjin Eye Institute, Tianjin, China
                Author notes

                Academic Editor: Leonardo Mastropasqua

                Author information
                http://orcid.org/0000-0003-0695-1651
                http://orcid.org/0000-0002-3387-1553
                http://orcid.org/0000-0002-0204-8361
                Article
                10.1155/2018/5461784
                6323435
                30671258
                abe935bd-5b9c-4888-992f-4b62f150ea3a
                Copyright © 2018 Shasha Yu et al.

                This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 30 April 2018
                : 19 September 2018
                : 11 November 2018
                Categories
                Research Article

                Ophthalmology & Optometry
                Ophthalmology & Optometry

                Comments

                Comment on this article

                scite_
                10
                0
                6
                0
                Smart Citations
                10
                0
                6
                0
                Citing PublicationsSupportingMentioningContrasting
                View Citations

                See how this article has been cited at scite.ai

                scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

                Similar content89

                Cited by3

                Most referenced authors252